<?xml version="1.0" encoding="UTF-8"?>
<?xml-stylesheet type="text/xsl" href="/oai-pmh.xsl"?>
<OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd">
  <responseDate>2026-09-21T15:26:32Z</responseDate>
  <request identifier="oai:www.ideals.illinois.edu:2142/26132" metadataPrefix="etdms" verb="GetRecord">https://www.ideals.illinois.edu/oai-pmh</request>
  <GetRecord>
    <record>
      <header>
        <identifier>oai:www.ideals.illinois.edu:2142/26132</identifier>
        <datestamp>2023-07-10</datestamp>
        <setSpec>col_2142_5131</setSpec>
        <setSpec>col_2142_8888</setSpec>
        <setSpec>com_2142_5130</setSpec>
        <setSpec>com_2142_8887</setSpec>
        <setSpec>com_2142_234</setSpec>
      </header>
      <metadata>
        <thesis xmlns="http://www.ndltd.org/standards/metadata/etdms/1.1/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:dc="http://purl.org/dc/elements/1.1/" xsi:schemaLocation="http://www.ndltd.org/standards/metadata/etdms/1.1/ http://www.ndltd.org/standards/metadata/etdms/1.1/etdms11.xsd http://purl.org/dc/elements/1.1/ http://www.ndltd.org/standards/metadata/etdms/1.1/etdmsdc.xsd">
          <dc:date>2011-08-25T22:15:30Z</dc:date>
          <dc:date>2011-08-25T22:15:30Z</dc:date>
          <dc:creator>Umeki, Robyn T.</dc:creator>
          <dc:contributor>Oelze, Michael L.</dc:contributor>
          <dc:date>2011-08-25T22:15:30Z</dc:date>
          <dc:date>2011-08</dc:date>
          <dc:description>A focused transducer will concentrate its signal energy at its focus, producing a clear
image of objects at that depth. However, spatial resolution and signal strength deteriorate away
from the transducer’s focus. Therefore, having only one focal length places a great restriction on
ultrasound imaging, especially with a single-element transducer. To extend the region of focus,
three techniques were examined. The first was the synthetic aperture focusing technique (SAFT),
a well-studied method that allows focus at every distance instead of just one distance. Using this
method, lateral resolution is restored and signal-to-noise ratio (SNR) is increased due to the
compounding of several scan lines. SAFT was combined with a virtual source technique to
further extend the region of focus. The second technique examined was coded excitation, a
means of improving SNR while maintaining the axial resolution. The third technique was tissue
harmonic imaging, which produces a narrower beamwidth and reduced sidelobes, and therefore
can improve spatial resolution and contrast of images.
Each technique is individually known to improve ultrasound image qualities, with its own
strengths and drawbacks. This study combines all three. In the final stage of this study using the
virtual source technique, harmonic imaging was implemented using coded excitation beyond the
transducer focus. Because of the low signal strength received under the synthetic aperture
technique, harmonic imaging has never been attempted using SAFT and a virtual source. There
has been no previous research to determine whether the techniques will build on each other or
detract from each other.
This study included simulations and experiments with the techniques applied to a single
scatterer, as well as experiments with a tissue-mimicking phantom. Implementing coded
excitation with SAFT led to a final SNR higher than was observed when applying SAFT alone.
Although coding helped to increase SNR, the sidelobes became much more visible.
Additionally, it was found that synthetic aperture was not a linear process and could cause
distortion when the transmitted pulse is very long. This problem was eliminated when time
compression was applied before SAFT. When applying tissue harmonic imaging, the sidelobes
were less prominent than before. However, the speckle increased and the SNR decreased, likely
due to the decreased signal strength of the transducer at the harmonic frequencies.
iii
This study has shown that synthetic aperture with a virtual source, coded excitation, and
tissue harmonic imaging can be combined to image beyond the focal length of a transducer.
However, there was no research on the maximum depth at which the technique can be practical.
Additionally, this study did not test the effectiveness of the techniques when applied to real
tissue.</dc:description>
          <dc:description>Item withdrawn by Mark Zulauf (zulauf@illinois.edu) on 2011-07-20T14:05:38Z
Item was in collections:
University of Illinois Theses &amp; Dissertations (ID: 1)
No. of bitstreams: 1
Umeki_Robyn.pdf: 740075 bytes, checksum: b72d6c93208a5a63c7f1c0230c486dcc (MD5)</dc:description>
          <dc:description>Made available in DSpace on 2011-08-25T22:15:30Z (GMT). No. of bitstreams: 2
Umeki_Robyn.pdf: 740075 bytes, checksum: b72d6c93208a5a63c7f1c0230c486dcc (MD5)
license.txt: 4060 bytes, checksum: 4bfad0015831d78359b83adad80f89a8 (MD5)</dc:description>
          <dc:identifier>http://hdl.handle.net/2142/26132</dc:identifier>
          <dc:language>en</dc:language>
          <dc:rights>Copyright 2011 Robyn T. Umeki</dc:rights>
          <dc:subject>ultrasound imaging</dc:subject>
          <dc:subject>synthetic aperture</dc:subject>
          <dc:subject>coded excitation</dc:subject>
          <dc:subject>tissue harmonic imaging</dc:subject>
          <dc:title>Applying synthetic aperture, coded excitation, and tissue harmonic imaging techniques to allow ultrasound imaging with a virtual source</dc:title>
          <degree>
            <department>Electrical &amp; Computer Eng</department>
            <departmentCode>1933</departmentCode>
            <discipline>Electrical &amp; Computer Engr</discipline>
            <disciplineCode>1200</disciplineCode>
            <grantor>University of Illinois at Urbana-Champaign</grantor>
            <level>Thesis</level>
            <name>M.S.</name>
            <program>MS:Electr &amp; Computer Eng-UIUC</program>
            <programCode>10KS1200MS</programCode>
          </degree>
        </thesis>
      </metadata>
    </record>
  </GetRecord>
</OAI-PMH>